Automated Printing of Electrochemical Immunoassay Microarrays: Studies of Conditions Affecting Alkaline Phosphatase Enzyme Label Activity

by Williams, C. R.; Fritsch, I.

Microarray printing of components of an enzyme-linked immunosorbant assay (ELISA) is attractive for constructing miniaturized platforms to simultaneously and quantitatively analyze multiple chemical species in clinical, environmental, and forensic applications. A truncated version of an electrochemical ELISA was used to address a major challenge of this approach, namely the effect of evaporation on assay activity. An antibody-alkaline phosphatase conjugate (Ab-AP) was printed with a microarrayer onto an N-hydroxysuccinimide-ester-terminated self-assembled monolayer on gold. The activity of the enzyme label AP to convert the substrate p-aminophenyl phosphate to electroactive p-aminophenol was followed as a function of different conditions via detection by cyclic voltammetry at microband electrodes. It was found that enzyme activity decreased by 60% from 30 min to 4 h after evaporation. However, printing in ambient humidity ("dry") as opposed to printing in a humidified chamber, resulted in the highest assay signals observed and low nonspecific adsorption.

Journal
ECS Transactions
Volume
28
Issue
21
Year
2010
Start Page
19-33
URL
https://dx.doi.org/10.1149/1.3496016
ISBN/ISSN
1938-6737; 1938-5862; 9781566778473
DOI
10.1149/1.3496016